Optium

MediSense®

OptiumTM

Blood Glucose Electrodes

Important: To ensure accurate results, you must read the MediSense®Optiumä User's Guide before testing.

Important!Information when changing to end-fill electrodes: Previously stored test results between 27.8 mmol/l or 500 mg/dL and 33.3 mmol/l or 600 mg/dL are now displayed as "HI". When using the result averaging function, your average should be slightly lower. As older results are removed from memory this problem will correct itself. We strongly recommend that you do not use our original electrode again.

If using Precision Link· Data Management Systems, please contact Customer Service, for additional information, BEFORE using these electrodes.

Intended Use

For use outside the body (in vitro diagnostic use). The MediSense Optium System quantitatively measures glucose (D-glucose) in fresh capillary whole blood (i.e., from the finger).

Important: Use only with MediSense Opium System. Not for use with any other MediSense Systems.

Test Principle

When the blood sample is applied to the electrode, the glucose in the blood reacts with the chemicals on the electrode, producing a small electrical current. This current is measured and then a result is displayed by the sensor. The size of the current depends on the amount of glucose in the blood sample.

Storage and Use

Precautions

Sample Collection and Preparation

Before obtaining a blood sample, use warm soapy water to wash your hands and dry completely. Warm water helps to increase the blood flow to your fingertips. If you have difficulty in obtaining a blood sample, hanging your arm down before lancing your finger may also help increase the blood flow. Avoid excessive squeezing of the puncture site. When the capillary blood sample is obtained, it must be used immediately.

Healthcare Professionals

Note: Capillary blood may be collected into heparin-containing or potassium EDTA-containing capillary tubes and used within 30 minutes. Do not use tubes containing fluoride or oxalate.

Procedure

Refer to the MediSense Optium User's Guide for detailed testing instructions.

Materials provided: MediSense Optium Blood Glucose Electrodes and Calibrator.

Materials required but not provided: MediSense Optium Sensor and User's Guide, Precision Controls, lancing device and lancets.

Blood Application


Glucose dehydrogenase (Microbial) > 0.05 U
NAD+ (as sodium salt) > 7.2 m g
Phenanthroline quinone > 0.4 m g
Non-reactive ingredients > 16.0 m g

 

Limitations of Procedure

Performance Characteristics

MediSense Optium Blood Glucose Electrodes are calibrated to give plasma-equivalent glucose values.

Assay Range: 1.1-27.8 mmol/l (20-500 mg/dL).

Test Time: 20 seconds

Precision

Precision testing shows that result typically vary by no more than 2.9% to 5.1%. Testing was performed within a laboratory using venous blood samples on the system.

 

Low

Mid-Low

Mid-High

High

Mean mmol/l

2.2

4.2

12.1

17.8

(mg/dL)

(40)

(76)

(218)

(320)

SD mmol/l

0.11

0.17

0.39

0.50

(mg/dL)

(2.0)

(3.0)

(7.0)

(9.0)

CV %

5.1

3.5

3.4

2.9

Accuracy

Accuracy testing shows that results compare well with a laboratory reference method and that results are comparable between trained operators and lay users. The accuracy of the system was assessed at three independent sites by comparison with the YSI Glucose Analyzer as used in clinical laboratories. The data was obtained using fresh capillary whole blood. The glucose range was 3.1-25.8 mmol/l (56-464 mg/dL).

YSI Samples

Plasma

No. of samples

167

Slope

0.90

Intercept mmol/l (mg/dL)

0.92 (16.6)

r (corr. coef.)

0.98

Reference

  1. Burtis, C.A. & Ashwood, E.R., Clinical Chemistry, Third Edition, Philadelphia, W.B. Saunders Co. (1999) p. 1815.

1 2 3 4

Top